Omron F250-UME Software Manual page 365

Table of Contents

Advertisement

EC Defect
2-15-5-1
Defect
Original image
Line defect
1
Set
measureme
nt
CHECK
Original image
Line defect
STEP 1: Setting Measurement Conditions
Measurement conditions
Extracting level :
Defect color
Defect width
Defect length
End
Black*, white
The asterisk (*) indicates the default setting.
This defect inspection operation detects line defects on a removal image.
Only the 2 parallel lines in the edge-extracted image are detected as defects.
Circle
removed
2
Set
inspection
region.
Any lines in the measurement object that are detected as only single lines,
such as the exterior, are ignored. In this case, it is not necessary to set
removal regions.
Set the conditions for detecting lines as defects.
Reduce this number if the lines to be detected as defects are
broken and detected as multiple lines. Increase this number if
defect lines are not found. (1 to 5, 3*)
Parallel lines are detected as defects. Set the edge widths of
3
defects in pixel units between 0 and 100 (1*:10*) so that two
quite separate lines are not detected as defects.
:
Black
: [
1:10]pix
: [
1:50]pix
Set the length of lines to be detected as defects
in pixel units. (0 to 999, 1:50*)
Removal image
Removal image
Lines detected as only single
lines are not detected as defects.
Section 2-15
Inspection region
Inspection region
Edge-extracted image
Defect width
Defect length
2-15-(32)

Advertisement

Table of Contents
loading

Table of Contents