VeEX MPA User Manual page 130

Multi-protocol analyzer
Table of Contents

Advertisement

MPA_e_manual_D07-00-129P_RevA00
Function
Settings Controls
Interface
Description
If Coupled is selected, settings changed on the Receive
screen will automatically change on the Transmit screen as
well.
If Independent is selected, settings established on the
Receive screen will only affect Receive mode.
Selects the receive Interface Rate for the MPM-400G, MPM-
100G or MPM-100AR card.
400G Interface rates require a CFP8 pluggable laser
module.
100G Interface rates require a CFP4 or QSFP28
pluggable laser module.
40G Interface rates require a QSFP+ pluggable laser
module.
32G and 25G Interface rates require a SFP28
pluggable laser module.
16G and 10G Interface rates require a SFP+
pluggable laser module.
Constant LOS or LOF alarms may be present when an
unsupported Interface rate is selected or if the required laser
module is unplugged.
The wavelength of the selected interface is determined by
the specifications of the laser module being used.
Verify that the inserted laser module supports the desired
Interface rate prior to selecting it.
Unframed BERT Interfaces are only valid when the
number of Physical Lanes equal the number of Optical
Lanes. Refer to the
MLD Receive - SIG
MLD Tab
screen.