Calibration of Internal Sensors

Available for CS20. Calibrate Disto tilt sensor is availableif the CS20 is equip-
ped with a DISTO.
Select Leica Captivate - Home: Settings\System\Calibrate internal sen-
sors or Calibrate Disto tilt sensor.
Calibrate the internal sensors prior to use when the compass functionality is
used.The calibration wizard guides through the calibration process.
Re-calibrate the DISTO tilt sensor if the CS20 has been dropped or heavily
bumped.
The calibration wizard guides through the calibration process.
Steps to check if a re-calibration is required:
1.
Place the CS20 on a flat surface about 20 m away from a wall but
aiming at the wall.
2.
Access a panel where it is possible to view the tilt value: Select Leica
Captivate - Home: Measure. Press Fn Tools. Select Measure hid-
den point.
3.
The tilt value must be close to 0.
4.
If the tilt differs from 0, then the DISTO calibrate the tilt sensor.