Leica Captivate Technical Reference Manual page 270

Table of Contents

Advertisement

Check & Adjust,
Step 2
Check & Adjust,
It is recommended to
repeat the last cali-
bration routine at
least three times
Check & Adjust,
Results
270
For MS60/TS60:
If Calibrate the telescope
camera has been checked, aim
at the same target accurately
with the telescope camera using
the view finder and the digital
crosshair on the display. Meas-
ure to measure and to continue
to the next panel.
Motorised instruments change
automatically to the other face.
Non-motorised instruments guide to the other face using the Tele-
scope Positioning panel.
2.
Measure to measure the same target in the other face and to calcu-
late the instrument errors.
If one or more errors are bigger than the predefined limits, the pro-
cedure must be repeated. All measurements of the current run are
rejected and are not averaged with the results from previous runs.
Key
Description
Next
To measure the target.
Description of fields
Field
Option
Number of
Display only
measure-
ments
All other
Display only
fields
Measure at least two more runs.
Next step
IF
more runs must be added
no more runs must be added
Select the instrument errors which you want to accept and then store.
Do not check a box to keep the currently used error active on the instrument
and rejects the new one.
180°
008372_001
Description
Shows the number of runs. One run consists
of a measurement in face I and II.
The standard deviations of the determined
adjustment errors are displayed. The stand-
ard deviations can be calculated from the
second run onwards.
THEN
select Add another calibration loop and
press Next.
select Finish the calibration & store the
results and press Next to accept the meas-
urements and to access the results panel.
Settings - TS instrument