Check & Adjust,
Make first tilt meas-
urement in any face
Check & Adjust,
Step 2
Check & Adjust,
It is recommended to
repeat the last cali-
bration routine at
least three times
274
levelled up using the electronic level
protected from direct sunlight
acclimatised to the ambient temperature, approximately 2 minutes per °C
difference compared to the storage place.
Key
Description
Next
To measure the target.
1.
Measure to measure the first
face. No target has to be aimed
at.
Motorised instruments change
to the other face and release a
measurement automatically.
2.
Non-motorised instruments guide to the other face using the Tele-
scope Positioning panel.
3.
Measure to release the measurement in the other face.
If one or more errors are bigger than the predefined limits, the pro-
cedure must be repeated. All measurements of the current run are
rejected and are not averaged with the results from previous runs.
Key
Description
Next
To measure the target.
Description of fields
Field
Option
Number of
Display only
measure-
ments
l Compo-
Display only
nent quality
(1 σ) and t
Component
quality (1 σ)
Measure at least two more runs.
Next step
IF
more runs must be added
180°
008374_001
Description
The number of runs. One run consists of a
measurement in face I and II.
The standard deviations of the determined
adjustment errors. The standard deviations
can be calculated from the second run
onwards.
THEN
select Add another calibration
loop and press Next.
Settings - TS instrument