Compensator Adjustment (L, T) - Leica Captivate Technical Reference Manual

Table of Contents

Advertisement

Check & Adjust,
It is recommended to
repeat the last cali-
bration routine at
least three times
Check & Adjust,
Results
22.7.6
Access
Description
Settings - TS instrument
Field
Option
a T-axis
Display only
quality (1 σ)
Measure at least two more runs.
Next step
IF
more runs must be added
no more runs must be added
Key
Description
Back
To return to the previous panel.
Finish
To accept and record the new determined tilting-axis error. If
the report recording has been enabled, then the results are
written to or appended to an existing report sheet.
Redo
To reject the result and to repeat the complete check and
adjust procedure.
Description of fields
Field
Option
New
Display only
Old
Display only

Compensator Adjustment (l, t)

In Check & Adjust, What do you want to do? select Check & adjust the
compensator and press Next.
The compensator adjustment procedure determines the following instrument
errors:
Instrument error
l
t
Before determining the compensator index errors, the instrument has to be:
Description
The standard deviation of the determined
tilting-axis error. The standard deviation can
be calculated from the second run onwards.
THEN
select Add another calibration
loop and press Next.
select Finish the calibration &
store the results and press Next to
accept the measurements and to
access the results panel.
Description
The new determined and averaged tilting-
axis error.
The old instrument error, which is valid on
the instrument.
Description
Compensator longitudinal index error
Compensator transversal index error
273