Check & Adjust; Overview; Tilt Sensor Calibration - Leica 3D Disto User Manual

Three-dimensional measuring and projection system
Hide thumbs Also See for 3D Disto:
Table of Contents

Advertisement

8
8.1
Description
Electronic adjustment
8.2
Description
Adjustment step-by-
step
Check & Adjust
Check & Adjust

Overview

Leica Geosystems instruments are manufactured, assembled and adjusted to
the best possible quality. Quick temperature changes, shock or stress can
cause deviations and decrease the instrument accuracy. It is therefore recom-
mended to check and adjust the instrument from time to time. This check and
adjust can be done in the field by running through specific measurement pro-
cedures. The procedures are guided and must be followed carefully and pre-
cisely as described in the following chapters. Some other instrument errors
and mechanical parts can be adjusted mechanically.
The following instrument errors can be checked and adjusted electronically:
Crosshairs offset
V-index
Tilt sensor
All calibration settings can also be reset to factory default.
During the manufacturing process, the instrument errors are carefully determ-
ined and set to zero. As mentioned previously, these errors can change and it
is highly recommended to redetermine them in the following situations:
After rough or long transportation
After long storage periods
If the temperature difference between the current environment and at the
last calibration is more than 20°C

Tilt Sensor Calibration

Before Calibration:
3D Disto_033
a
Tilt sensor offset
Step
Description
This calibration is done automatically. You must only ensure that
the 3D Disto is set up < 3°.
1.
Calibration is started in Menu » Device » Calibration.
After Calibration:
a
3D Disto_034
77