Configuring Check & Adjust; Adjusting The Circular Level Of The Instrument And Tribrach - Leica Captivate Technical Reference Manual

Table of Contents

Advertisement

22.7.8
Access
Check & Adjust
22.7.9
Adjusting the Circular
Level Step-by-Step
276
Configuring Check & adjust
In Check & Adjust, What do you want to do? select Check & adjust set-
tings and press Next.
Key
Description
Next
To accept changes and to continue with the subsequent
panel within the wizard.
Back
To return to the previous panel.
Description of options
Option
Description
Every 2
If one or more adjustment values were determined longer
weeks,
ago than the time specified with this parameter, then a
Once a
reminder message is displayed each time the instrument is
month,
turned on. The reminder helps to redetermine the instrument
Every 3
errors regularly.
months,
Twice a year
or
Once a year
Never
A reminder message to readjust the instrument is never dis-
played. This setting is not recommended.
Next step
Next to change to the Report sheet panel.

Adjusting the Circular Level of the Instrument and Tribrach

4
5b
5b
5b
008375_001
1.
Place and secure the instrument into the tribrach and onto a tripod.
2.
Using the tribrach footscrews, level the instrument with the elec-
tronic level.
3.
Select Settings\TS instrument\Level & compensator to access
the Level & Compensator panel.
4
1
2
Settings - TS instrument
4
5b
5b
5b