Select the "_Vascular" or "_Vascular+IGS" preset (2).
X
When choosing this preset profile, a popup will appear to notify the
user about the profile limitations.
It is highly recommended to change an existing surgeon profile or
add a new surgeon profile to integrate the GLOW800 function in a
full workflow.
The following functions are assigned when selecting the GLOW800
preset:
Left handle
no function
GLOW800
Play loop
Prior loop
You can switch back to white light mode by pushing the
joystick of the left handle towards the left again.
2
Right handle
Lamp+
Record video
Lamp-
GLOW800 / Ref. 10 748 652 / Version 01
9
Check and adjust the
illumination and functions
9.1
Pre-operation checklist
(GLOW800)
Cleaning optical accessories
Check optical accessories for cleanliness.
X
Remove dust and dirt.
X
GLOW800 application
When using GLOW800 please ensure to have a Doppler
X
Ultrasound or similar device in place, in case of none or
insufficient blood flow visualization out of the ICG/GLOW800
procedure is given.
Balancing
Balance the microscope after refitting (see user manual of
X
ARveo 8).
If you plan to use stereoscopic visualization, ensure that a suitable
visualization device is available. Ensure that the specific precautions
for stereoscopic visualization are taken care of.
Operational check
Switch on the microscope.
X
Switch on the illumination.
X
Check the microscope illumination.
X
Test the GLOW800 with the test card.
X
Sterility
Apply the sterile drape.
X
For sterilizable components of the ARveo 8, refer to the
corresponding user manual.
Take
picture
Check and adjust the illumination and functions
25