46
are corrected twice: first by the determined ATR errors for Hz and V, and then
by the individual small deviations of the current aiming.
•
Adjusting a combination of instrument errors
Refer to "5.3 Combined Adjustment (l, t, i, c and ATR)".
Adjusting the circular level
•
Refer to "5.4 Adjusting the Circular Level of the Instrument and Tribrach".
Adjusting the laser/optical plummet
•
Refer to "5.6 Inspecting the Laser Plummet of the Instrument".
Adjusting the tripod
•
Refer to "5.7 Servicing the Tripod".
Combined Adjustment
(l, t, i, c and ATR)
The combined adjustment procedure determines the following instrument
errors in one process:
Instrument error
l, t
i
c
ATR Hz
ATR V
The following description explains the most common settings:
1.
Connect the device with the Instrument.
2.
Select Devices from the Home Menu.
3.
•
Select your instrument and tap the arrow.
Select Sensor Calibration.
•
☞
It is recommended to use a clean Leica circular prism as target. Do
not use a 360° prism.
4.
Select the incl. ATR Calibration option if you like to calibrate
•
the ATR.
•
Tap
Follow the wizard which guides through the calibration.
•
Description
Compensator longitudinal and transversal index
errors
Vertical index error, related to the standing axis
Horizontal collimation error, also called line of
sight error
ATR zero point error for horizontal angle option
ATR zero point error for vertical angle option
to see which instrument errors are determined.
Check & Adjust