GE Senographe DS Manual page 734

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Job Card ELE A043 - AWS Monitor Calibration
7-4
Display the SMPTE Test Pattern
Display the SMPTE pattern on the AWS Monitor by using the following steps:
1. From the Browser, launch the Service Desktop (see
2. From the Service Desktop, click the Calibration button
then click the START button to display the SMPTE Test Pattern.
Central black level square
used to do the monitor bright-
ness adjustment.
3. Ensure that the mouse cursor is in the image of the SMPTE Test Pattern so that is correctly dis-
played.
Note:
Do NOT use the "FFDM pattern" that is accessible from the browser patient list.
4. Once the SMPTE Test Pattern is displayed, go to
7-5
Move the OSD to a Convenient Position on 18" Monitors
It may be desirable to move the OSD display toward the upper part of the screen to fully observe the fol-
lowing squares of the SMPTE pattern:
central white level square
central black level square
5% / 0% grey level square
Chapter 7
Service Information and Procedures Class A 2385072-16-8EN
Launching the Service Desktop on page
and select Peripherals/AWS monitor,
Central white level square
used to do the monitor back-
light adjustment, and contrast
adjustment.
section
7-5.
Page no. 734
Senographe DS
120).
JC-ELE-A-043.fm