Revision 1
Job Card ELE A020 - Declare and Set Printers (Filming Devices)
7-2
Agfa Printers
7-2-1
Agfa Drystar AXYS DICOM Printer
DICOM printer label: Any name you wish to see appearing on the Film Composer e.g., Agfa AXYS.
Hostname: Use printer Hostname if any, or a name unique to your Network e.g., DSAXYSM.
Network address: A unique IP address given by the hospital Network Administrator, e.g.,
3.48.16.136.
Application Entity Title (AET):Default value is AXYS_M , however contact the Agfa engineer for
Port number:
Layouts:
Film size:
Pixel size
Configuration information:Must be PERCEPTION_LUT=OEM919
Density
Magnification type:
Smoothing factor:
Trim:
Printer Memory Size
Chapter 7
Service Information and Procedures Class A 2385072-16-8EN
the correct AET value to use
Must be 104
Must be 1 x 1
, 1 x 2
Must be 8 in x 10 in W = 3828
10 in x 12 in W = 4892
50
Must be Min: no value Max: 370
Must be Cubic
Must be 143
Must be NO
Must be 40