GE Senographe DS Manual page 304

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Cable Lay-out and Pin-out
1-26. W127 - EMC Filter
EMC filter E101 to Terminal block ST101.
1-27. W129 - Terminal Block Interconnection
Terminal Block Interconnection from X4 P' to B2/1.
1-28. W201 - Lift Optical Fork Connect (Upper)
Lift Board PL201 J4 to opto detector S201:
Designation
VCC_Led_1
GND
Fork_Opt_Out_
1
GND
1-29. W202 - Lift Optical Fork Connect (Lower)
Lift Board PL201 J5 to opto detector S202.
1-30. W203 - Lift Potentiometer
Lift Board PL201 J11 to potentiometer R201:
Designation
VCC
Pot.
GND
1-31. W204 - LCD Connect
Lift Board PL201 J15 to LCD DS201.
Chapter 3
Type
Level
VCC
5 V
Ground
0 V
Open coll.
0 / +5 V
Ground
0 V
Type
Level
VCC
5 V
Analog.
0 / 5 V
Ground
0 V
Service Information and Procedures Class A 2385072-16-8EN
Pin N°
J4-1
J4-2
J4-3
J4-4
R201
Pin N°
J11-1
J11-2
J11-3
Page no. 304
S201
W 201
2345937
300
A
K
C
E
Vcc
Vout
GND
Senographe DS
J4
1
2
3
4
J11
1
2
3
Cables.FM