GE Senographe DS Manual page 1230

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Job Card D/R A222 - Clutch and Cable
The cables from the clutch to the Rotation Board PL202 must be secured as shown:
Pass a cable tie (A) through the
hole in the curtain metal support
to secure the cables behind the
curtain.
The free length between the two
cable ties (A and B) is approx. 60 cm
(24 inch).
Use some adhesive tape to secure
the two cables together in this area.
Do not use cable ties; they can
catch on other items during
movement.
6-4
Check Rotation Movement
Move the compression paddle out of compression. Then check that all four sets of keypad control but-
tons on the arm and the tube head are operational (check each button for arm rotation movement).
7
COMPLETION
None
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
When installing the covers, place these cables inside the
lower cover with the keypad cable. The cables must
move freely; check that they cannot catch on anything.
Page no. 1230
Senographe DS
Use a cable tie (B) to
secure the cables at
this point.
JC-DR-A-222.fm