leave the other end connected to the framework.
15. Remove the four nuts (16) (8 mm socket and torque wrench) that secure the defective
AC/DC Module to the framework, and remove the defective AC/DC Module.
16
JC-DR-A-194.fm
Service Information and Procedures Class A 2385072-16-8EN
15
AC/DC Module
Page no. 1157
Senographe DS
Job Card D/R A194 - AC/DC Module
16