GE Healthcare
Revision 1
Central Listing
4-2. Stepper Board
•
Board-Compression PL303; see component index page 157
•
Board-Tilt PL401; see component index page 158
DS5
DS7
J5
LED
Function
DS1
DS2
DS3
Synchro Grid
DS4
DS5
DS6
RT1 Valid Line
DS7
Compression Brake Actived
(does not exist for tilt board)
DS8
DS9
DS10
Microcontoller heartbeat
Chapter 3
J1
DS1
DS9
J3
J11
J12
48 V
3,3 VA
Not Used
5 V
RX-CAN
3,3 V
Service Information and Procedures Class A 2385072-16-8EN
.
J7
J6
DS2
J21
J8
Color
Green
ON when the 48 V DC supply voltage is OK
OFF when the 48 V DC supply voltage is defective
Green
ON when the 3.3 VA analog supply voltage is OK
OFF when the 3.3 VA analog supply voltage is defective
Yellow
Yellow
N/A
Green
ON when the 5 V supply voltage is OK
OFF when the 5 V supply voltage is defective
Yellow
ON when the valid line is active (movement is enabled)
OFF when the valid line is inactive (movement is not
enabled)
Green
ON when the rotation brake is inactive (rotation move-
ment is enabled)
OFF when the rotation brake is active (rotation move-
ment is not enabled)
Yellow
ON when CAN data (or command) received.
OFF when no CAN data (or command) received.
Green
ON when the 3.3 V supply voltage is OK
OFF when the 3.3 V supply voltage is defective
Red
Blinks when the HC12 microcontroller is alive, represent-
ing the heartbeat of the HC12 microcontroller.
OFF when the HC12 microcontroller is dead.
Page no. 288
.
F1
J10
DS4
DS3
DS6
DS8
J22
J2
DS10
F3
J4
J9
Behavior
Senographe DS
Indicators and Switches.fm