Job Card DIAG A1000 - Gantry Diagnostic
132/009 error code (CAN code is 0x501C)
132/010 error code (CAN code is 0x501D)
132/025 error code (CAN code is 0x501E)
132/026 error code (CAN code is 0x501F)
132/027 error code (CAN code is 0x5020)
Ensure that nothing is pressing on any keypad button, then reboot the Gantry.
If the same message appears:
-
Check the keypad cable and connections. If the cable is faulty, change it.
-
Disconnect the keypad and connect another of the installed keypads in its place, then reboot the
Gantry. If the keypad is found to be faulty, change it.
-
If the error persists, change the Tube Tilt Board (see
1365).
FRUs:
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Tube head keypad PL404
Tube Tilt Board
W407 cable for keypad
Page no. 964
Job Card D/R A266 - Tube Tilt Board on page
Senographe DS
JC-DIAG-A-1000.fm