Scenario Pre A002 - Pre-Purchase Site Visit - GE Senographe DS Manual

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1

Scenario PRE A002 - Pre-purchase Site Visit

1
CONTEXT
This scenario provides a check list for use in planning and carrying out a pre-purchase site visit.
2
STEERING GUIDE
#
Requirement
Altitude
1
Check that product specifications are
compatible with the altitude of the site
Operating conditions
2
Check that operating the temperature
and humidity requirements can be met
Rom layout
3
Check that an adequate room is avail-
able, with suitable floor and access
Electrical supply
4
Check availability of suitable supply
Networking
5
Check possibility of connection to hos-
pital network
Insite connection
6
Check availability and type of broad-
band connection
SC-PRE-A002.fm
Service Information and Procedures Class A 2385072-16-8EN
Scenario PRE A002 - Pre-purchase Site Visit
Reference
Chapter
4 Pre-Installation
System
Requirements,
section
1 Environmental
Requirements
Chapter
4 Pre-Installation
System
Requirements,
section
1 Environmental
Requirements
Chapter
4 Pre-Installation
System Requirements
Chapter
4 Pre-Installation
System
Requirements,
section
5 Electrical
Requirements
Chapter
4 Pre-Installation
System
Requirements,
section
12 Networking
Connections
Chapter
4 Pre-Installation
System
Requirements,
section
11 Insite Connec-
tion
Page no. 365
Senographe DS
Chapter 5
Who?
Hospital engi-
neer
Heating engineer
Site planner
Hospital engi-
neer
Hospital engi-
neer, GEMSE
HMS representa-
tive
Hospital engi-
neer
Done?