GE Senographe DS Manual page 933

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Code
Category
181/006
HALT
181/007
HALT
181/008
HALT
181/009
HALT
181/010
HALT
ERR-SUB-A-003.fm
Text in Log File or
on LCD
STEREOTAXY
HARDWARE FAIL-
URE
STEREOTAXY
HARDWARE FAIL-
URE
STEREOTAXY NEE-
DLE PATH FAILURE
STEREOTAXY
FAILURE
STEREOTAXY
SOFTWARE FAIL-
URE
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
Hardware failure with the stereotix power.
1. Change the CPU Board on the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
2. Perform a Calibration of the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
Hardware engine failure with the stereotix A/D con-
verter.
1. Change the CPU Board on the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
2. Perform a Calibration of the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
Security Failure: Needle path failure.
Try to reproduce the defect by playing with the nee-
dle path lock button.
• If defect is reproducible, the Z-axis trolley is faulty
and you must change the Stereotactic Positioner
(see the Stereotaxy Service Manual).
• If defect it is not reproducible, the faulty FRU is the
Stereotactic Positioner CPU board, and you must
change the CPU Board on the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
Security Failure: Power switch failure.
1. Change the CPU Board on the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
2. Perform a Calibration of the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
Software failure with the stereotix sequencing.
1. Unplug and replug the stereotix then reboot the
system from the X-Ray Console.
2. If the problem persists, replace the stereotix unit.
Page no. 933
ERR SUB A003 - Gantry Error Messages
Senographe DS
Job Card
Chapter 9