GE Healthcare
Revision 1
Pre-Installation System Requirements
1.

ENVIRONMENTAL REQUIREMENTS

1-1. Atmospheric Pressure Limits (Storage and Operational)
(operation, storage and transportation*)
* - During transportation a pressurized environment must be used to maintain the atmospheric
pressure limits.
1-2. Preventing Thermal Shock to the Detector
To prevent thermal shock to the detector, the ambient tempertaure gradient must not exceed the values
as specified below.
Temperature range
Ambient temperature
gradient must not
exceed
1-3. Storage Requirements - temperature and humidity
1-3-1.
General requirements
-
The system includes a detector assembly in its casing, which is sensitive to changes in tempera-
ture and humidity, and is water-cooled when in use.
-
The specified storage requirements assume that the all the equipment remains in its packaging,
including the protection for the detector.
1-3-2.
Before installation - short term
For short-term storage (less than 5 days), refer to the storage requirements table below.
1-3-3.
Before installation - long term
For long-term storage (more than 5 days), it is recommended that the detector assembly is kept in an
area with relatively dry and low temperature, or area with air conditioning. For example, relative humidity
less than 50%.
Chapter 4
Atmospheric pressure
Min.
700 hPa
Less than 0°C
Less than 32°F
4.5°C per hour
40.1°F per hour
Relative humidity
(non-condensing)
Min.
Max.
5%
95%
Service Information and Procedures Class A 2385072-16-8EN
Altitude (from sea level)
Max.
1060 hPa
0°C to 40°C
32° to 104°F
1.5°C per minute
34.7°F per minute
Temperature
Min.
–20°C
–4ºF
Page no. 324
Min.
Max.
0 m
3000 m
0 ft
9840 ft
Greater than 40°C
Greater than 104°F
4.5°C per hour
40.1°F per hour
Max.
50°C
122ºF
Senographe DS
Chap-Pre-Requirements.fm

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents

Save PDF