Job Card ELE A043 - AWS Monitor Calibration

1
SUPPLIES
None
2
TOOLS
A calibrated Luminance Meter that permits measuring the luminance of the monitor in contact (Direct)
mode as illustrated below.
Remove canon and place foam ring
Luminance Meter
Siemens Serial Spot Meter
configured in Direct mode.
The Away/Direct button toggles between away and direct modes. To select the Direct mode, keep
pressing the Away/Direct button until D is displayed on the lower left corner of the Spot Meter
screen.
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 40 minutes
JC-ELE-A-043.fm
Service Information and Procedures Class A 2385072-16-8EN
(P/N 2319223) must be used with the foam ring attached, and must be
Foam Ring
Direct mode selected
Away/Direct mode
button selector