Code
Category
110/001
HALT
110/006
HALT
110/007
HALT
ERR-SUB-A-003.fm
Text in Log File or
on LCD
ROTATION RELAY
DEFECT
ARM TURNED
RIGHT TOO FAR
ARM TURNED LEFT
TOO FAR
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
There is a relay defect at power-up. A relay (on the
rotation board) is mounted in parallel with the H-
bridge of the motor. This relay, when shut, creates a
short-cut of the H-bridge. At power-up, the node
does the following:
- injects 15 V to the H-Bridge
- shuts the relay and checks that there is 0 V
- re-opens the relay and checks that there is 15 V. If
not, this signal is produced.
1. Reboot the system. Check the error logs to see
whether the 110/001 ROTATION RELAY
DEFECT error occurs within the same time as
another POSITION ERROR (e.g. rotation or
compression).
2. If a POSITION ERROR (e.g. rotation or com-
pression) does exist within the same time, ignore
the 110/001 ROTATION RELAY DEFECT error.
Do not change the Rotation Board, and take
appropriate action to resolve the POSITION
ERROR.
3. If a POSITION ERROR (e.g. rotation or com-
pression) does not exist within the same time,
the 110/001ROTATION RELAY DEFECT error is
valid. Change the Rotation Board and then re-
calibrate the Rotation Arm.
The Arm is over the software specified limit for the
farmost right position. Occurs when the Arm position
is not between the software limits at power-up
(including a tolerance to be defined). Typically, this
can happen when Arm has been moved manually.
Shut down the Gantry, move lift down manually and
reboot the Gantry.
The Arm is over the software specified limit for the
farmost left position. Occurs when the Arm position
is not between the software limits at power-up
(including a tolerance to be defined). Typically, this
can happen when Arm has been moved manually.
Shut down the Gantry, move lift up manually and
reboot the Gantry.
Page no. 865
ERR SUB A003 - Gantry Error Messages
Senographe DS
Job Card
D/R A221
page
1223
CAL A021
page
1679
Chapter 9
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.