Advertisement

GE Healthcare
Revision 1
Central Listing
2-3-5.
Generator Command Board 400PL1 Test Points
Test Point
TP1
TP2
TP3
TP4
TP5
TP6
TP7
TP8
TP9
TP10
TP11
TP12
TP13
TP14
TP15
TP16
TP17
TP18
TP19
TP20
TP21
TP22
TP23
TP24
TP25
TP26
TP27
TP28
TP29
TP30
TP31
TP32
TP33
TP34
TP35
TP36
TP37
TP38
Chapter 3
Service Information and Procedures Class A 2385072-16-8EN
Label (in French)
Surcharge_Filament
Measure_Phase_1
Measure_Phase_3
ICH_Carre
Consigne_ICH_Carre
Output B13/8
Input B8 (auto_regul_ch)
Measure_Phases
Reset ADR
C_T6
B_T6
C_T5
B_T5
C_T4
B_T4
C_T3
B_T3
C_T2
B_T2
C_T1
B_T1
Output B13/6
Clock B6
SEL_LC
Auto_Regul_Ch
Measure_Phase_TTL
Output_B16 (ICH_OND)
Fil_Coupé
Surcharge_Detectee
Marche_Chauffage (dealye)
Output B24/10
Presence_Chauffage
Entretien LC
Input B38/5
5000_LC
Output B35/10
Fin_Tour
Detect_SurInt_LC
Page no. 256
Function (in English)
Filament overload
RMS Health Current
RMS Health Current Reference
Stator Command Signals
Filament Cut
Overload Detected in Filament
Heater Working (delay)
Heater Presence
Normal RPM Speed
5000 RPM Speed
Tour Finish
Overload Current LC detected
Senographe DS