GE Healthcare
Revision 1
Code
Message text
057/001
061/000
062/000
063/006
066/000
066/001
066/002
066/003
067/000
068/xxx
071/081 E12 Checksum error
(gen)
ERR-SUB-A-001.fm
Service Information and Procedures Class A 2385072-16-8EN
Description
Focus: focus change request not trans-
mitted or focus fault.
The probable causes of the 057/001
error are that the:
• HV cable is interacting with the tube
framework
• tilt movement is restricted
• Filter Wheel movement is restricted
• Tube Tilt Mechanism is faulty
Gene. to Gantry communication driver:
no reply from Gantry after two attempts.
Gene. to control panel communication
driver: no reply from control panel after
two attempts
Gene. init tests: no Gantry status feed-
back after 10 requests
Gene date/time: access timeout during
date init
Gene date/time: access timeout during
read sequence
Gene date/time: access timeout during
write sequence
Gene date/time: access timeout during
time write sequence
Generator: Zero divide
Generator: floating point calculation
error. xxx is VRTX task number
Generator: checksum error on battery
saved memory
Page no. 839
ERR SUB A001 - Generator Error Messages
Recommended actions
Refer to
JC DIAG A1302 TUBE HEAD
CONTROL FAILURE on page 965
fix the problem.
No Poseidon boot. If the problem per-
sists, change the Gantry CPU Board.
If this error occurs in conjunction with
the error 063/006 then the Bucky grid
has failed to home. In this case you
must:
-
Remove the Bucky
-
Reset the Senographe
-
Reconnect the Bucky
No Poseidon boot. If the problem per-
sists, change the Gantry CPU Board.
If this error occurs in conjunction with
the error 061/000 then the Bucky grid
has failed to home. In this case you
must:
-
Remove the Bucky
-
Reset the Senographe
-
Reconnect the Bucky
Senographe DS
to
Chapter 9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents

Save PDF