GE Healthcare
Revision 1
ERR SUB A003 - Gantry Error Messages
Code
Category
102/041
WARNING
102/042
WARNING
102/043
DEBUG_O
NLY
102/044
DEBUG_O
NLY
102/045
HALT
102/046
DEBUG_O
NLY
Chapter 9
Text in Log File or
on LCD
VALID LINE WAS
ACTIVE ON LIFT
NODE
LIFT WAS
STOPPED WITH-
OUT STOP COM-
MAND
LIFT INVALID
BRAKE DEFECT
ERROR
LIFT INVALID
CLUTCH MAL-
FUNCTION ERROR
LIFT INVALID USER
INTERFACE
ERROR
LIFT INVALID COL-
LIMATOR BUTTON
ERROR
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
The valid motion line is active at power-up.
1. Change the Gantry CPU Board (see
R A190 - Gantry CPU Board on page
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration on
page
1683).
3. Perform a Collimation Check (see
A028 - Collimation Checks on page
The lift node stopped the motion by itself because
the button/pedal responsible for the motion was
released and no stop order was received before the
timeout.
1. Change the Gantry CPU Board (see
R A190 - Gantry CPU Board on page
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration on
page
1683).
3. Perform a Collimation Check (see
A028 - Collimation Checks on page
There has been a brake malfunction. The Lift node
normally does not report this error in any case.
Ignore this error. If the problem persists, contact your
OLC.
There has been a clutch malfunction. The Lift node
normally does not report this error in any case.
Ignore this error. If the problem persists, contact your
OLC.
One of the Gantry footswitch pedals have failed.
Press very slowly or very quickly on each Gantry
footswitch pedals until the 102/045 error is triggered
again to identify which pedal is defective.
Replace the
left
and/or
pedals.
There is a collimator button error.
Ignore this error. If the problem persists, contact your
OLC.
Page no. 862
Job Card D/
1131).
Job Card
Job Card ELE
679).
Job Card D/
1131).
Job Card
Job Card ELE
679).
right
faulty Gantry footswitch
Senographe DS
Job Card
D/R A190
page
1131
CAL A029
page
1683
ELE A028
page 679
D/R A190
page
1131
CAL A029
page
1683
ELE A028
page 679
ERR-SUB-A-003.fm