GE Healthcare
Revision 1
Job Card ELE A010 - Set Lift Travel Limits
The numbered circles 1, 2, 3 and 4 correspond to the procedural sections within this Job Card as follows:
1. 6-2-1,
Upper Optical Fork Positioning
2. 6-2-2,
Lower Limit Adjustment (Lift Potentiometer Lower Reference Voltage at 0 mm)
3. 6-2-3,
Upper Limit Adjustment (Lift Potentiometer Upper Reference Voltage at 650 mm)
4. 6-2-4,
Selection of the Lift Travel Limit (Completion)
The top of the Senographe system Tube Head can travel up to three discrete maximum heights, corre-
sponding to each of the three different lift travel limits. The maximum Tube Head heights and minimum
required ceiling corresponding to each of the three different Lift Travel Limits, are as follows.
Lift Travel Limit
650 mm
750 mm
850 mm (default setting)
* - Assumes that the lower lift travel limit is accurately set.
** - The recommended minimum ceiling height assumes that there is a marginal gap of 70 mm between the top of
the Tube Head and the ceiling.
Note:
The information in the table above assumes that you have accurately set the lower lift travel limit
to 670 ±2 mm from the floor.
Goal of Calibration:
•
Select the upper optical fork position to avoid collision with the ceiling.
•
Calibrate the Lift Potentiometer.
•
Set the lift lower limit and the lift travel limit in software, defining the minimum and maximum usable
heights.
Chapter 7
Service Information and Procedures Class A 2385072-16-8EN
Corresponding Upper
Optical Fork Position
Bottom
Middle
Top (default setting)
Page no. 566
Corresponding Max.
Tube Head Height *
2230 mm
2330 mm
2430 mm
Senographe DS
Recommended Min.
Ceiling Height **
2300 mm
2400 mm
2500 mm
JC-ELE-A-010.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated