Network Configuration for V4 Control Stations
For V4 Control Stations, proceed as follows to complete the network configuration.
1. At the Network Connectivity screen, ensure that Yes is selected for the Networked option, and press
the <F2> key to continue.
N e t w o r k C o n n e c t i v i t y
S p e c i f y Y e s i f t h e s y s t e m i s c o n n e c t e d t o . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
N e t w o r k e d
‐
‐
X
X
2. At the Configure Multiple Network Interface screen, ensure that only e100g0 is selected from the list
of Network Interfaces. Press the <Return> key to confirm the selection, and then press the <F2> key
to continue.
C o n f i g u r e M u l t i p l e N e t w o r k I n t e r f a c e s
M u l t i p l e n e t w o r k i n t e r f a c e s h a v e b e e n . . . . . . . . . . . . . . . . . .
N o t e : Y o u m u s t c h o o s e a t l e a s t o n e i n t e r f a c e t o c o n f i g u r e .
N e t w o r k I n t e r f a c e s
‐
‐
X
X
3. At the DHCP for e1000g0 screen ensure, that No is selected for the Use DHCP for e1000g0 option,
and press the <F2> key to continue.
DHCP for e1000g0
Specify whether or not ...
JC-ELE-A-021.fm
‐
‐
‐
‐
‐
‐
‐
‐
‐
Y e s
N o
‐
‐
‐
‐
‐
‐
‐
‐
‐
e 1 0 0 0 g 0
b c m e 0
Use DHCP for e 1000g 0
‐
‐
‐
‐
‐
‐
‐
Yes
X
X
No
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A021 - Networking Connection and Configuration
‐
‐
‐
‐
‐
‐
‐
‐
‐
‐
‐
Page no. 643