GE Senographe DS Manual page 988

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Job Card DIAG A002 - AWS Ethernet Port Diagnostics
3
AWS-IDC LINE TESTS
3-1
Internal Test (for V1/V2 Control Stations)
Open boot command
Type of test:
Note:
The internal test is only applicable to V1/V2 Control Stations.
1. Go to the ok prompt (see
2. At the ok prompt, enter:
test net
3. Relaunch the AWS application (see
Result interpretation:
After the command has been entered, the following information appears:
Internal loopback test –
Transceiver check –
where Result 1 and Result 2 are the results of each sub-test.
-
If Result 1 is not Succeeded, the AWS motherboard is faulty. You must replace the ADS Com-
puter (see
Job Card D/R A311 - AWS Unit on page
-
If Result 1 is OK & Result 2 is not ok, the AWS transceiver is faulty. You must replace the ADS
Computer (see
-
If Result 1 & Result 2 are ok, the AWS motherboard is good, and either the cable is faulty or the
IDC is faulty. In this case, replace the Ethernet cable between the ADS Computer and IDC with a
cable that is known to work. Then attempt to ping the IDC, by entering
ping idc
If the IDC responds to the ping command, the cable was faulty. If the IDC still does not respond to
the ping command then the IDC is faulty and needs replacing.
3-2
External Test (for All Control Stations)
SUN VTS
Type of test:
1. From behind the Browser, launch a command window (see
page
120).
2. Run the external test as follows:
For V1/V2 Control Stations, at the prompt in the command window, enter:
/opt/SUNWvts/bin/nettest –v -f -o dev=ce0,target=IDC
For V3 Control Stations, at the prompt in the command window, enter:
/opt/SUNWvts/bin/nettest –v -f -o dev=nge1,target=IDC
For V4 Control Stations, at the prompt in the command window, enter:
/opt/SUNWvts/bin/nettest –v -f -o dev=e1000g0,target=IDC
Chapter 9
Go to ok Prompt on page
Relaunch the AWS Application on page
Result 1
Result 2
Job Card D/R A311 - AWS Unit on page
Service Information and Procedures Class A 2385072-16-8EN
987).
1401).
1401).
Launching a Command Line Window on
Page no. 988
Senographe DS
987).
JC-DIAG-A-002.fm