Job Card DIAG A002 - AWS Ethernet Port Diagnostics
3
AWS-IDC LINE TESTS
3-1
Internal Test (for V1/V2 Control Stations)
Open boot command
Type of test:
Note:
The internal test is only applicable to V1/V2 Control Stations.
1. Go to the ok prompt (see
2. At the ok prompt, enter:
test net
3. Relaunch the AWS application (see
Result interpretation:
After the command has been entered, the following information appears:
Internal loopback test –
Transceiver check –
where Result 1 and Result 2 are the results of each sub-test.
-
If Result 1 is not Succeeded, the AWS motherboard is faulty. You must replace the ADS Com-
puter (see
Job Card D/R A311 - AWS Unit on page
-
If Result 1 is OK & Result 2 is not ok, the AWS transceiver is faulty. You must replace the ADS
Computer (see
-
If Result 1 & Result 2 are ok, the AWS motherboard is good, and either the cable is faulty or the
IDC is faulty. In this case, replace the Ethernet cable between the ADS Computer and IDC with a
cable that is known to work. Then attempt to ping the IDC, by entering
ping idc
If the IDC responds to the ping command, the cable was faulty. If the IDC still does not respond to
the ping command then the IDC is faulty and needs replacing.
3-2
External Test (for All Control Stations)
SUN VTS
Type of test:
1. From behind the Browser, launch a command window (see
page
120).
2. Run the external test as follows:
For V1/V2 Control Stations, at the prompt in the command window, enter:
/opt/SUNWvts/bin/nettest –v -f -o dev=ce0,target=IDC
For V3 Control Stations, at the prompt in the command window, enter:
/opt/SUNWvts/bin/nettest –v -f -o dev=nge1,target=IDC
For V4 Control Stations, at the prompt in the command window, enter:
/opt/SUNWvts/bin/nettest –v -f -o dev=e1000g0,target=IDC
Chapter 9
Go to ok Prompt on page
Relaunch the AWS Application on page
Result 1
Result 2
Job Card D/R A311 - AWS Unit on page
Service Information and Procedures Class A 2385072-16-8EN
987).
1401).
1401).
Launching a Command Line Window on
Page no. 988
Senographe DS
987).
JC-DIAG-A-002.fm

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Subscribe to Our Youtube Channel

Table of Contents

Save PDF