Job Card PHY A042 - Remove/Reinstall Generator Covers
b. Lower the cover to the floor so the holes (5) in the bottom end of the cover pass past the rods (6)
on each side.
c. Disconnect the ground wire (7) at the Generator end to avoid risk of creating a short circuit.
d. Fully remove the cover and store it in a safe place.
6. For the left and right covers, do the following:
a. Remove the two bolts (8) that secure the top of the cover (7 mm open ended wrench)and pull it
towards yourself.
Chapter 6
Service Information and Procedures Class A 2385072-16-8EN
6
5
7
8
Page no. 516
8
Senographe DS
JC-PHY-A-042.fm