Job Card ELE A067 - Repeat and Reject Analysis Configuration

1
SUPPLIES
AWS Application Software CD-ROM
2
TOOLS
None
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 10 minutes
4
SAFETY PRECAUTIONS
No specific safety precautions are applicable.
5
PREREQUISITES
The Repeat and Reject Analysis feature must have been installed, and the installation step Reset
Repeat & Reject Analysis OK must have been performed. In the booklet supplied with the AWS Applica-
tion Software CD, refer to the last step of Install from Scratch or Update from CD-ROM.
6
PROCEDURE
1. Discuss the required configuration of the Repeat and Reject Analysis feature with the customer.
2. Before activating the feature, ensure that all names held on the system for technologists who are
concerned with this feature are correctly entered.
! Notice:
Check that no name includes unwanted blank characters. Data for any technologists whose name
contains two or more consecutive blank characters are not taken into account for the analysis.
3. Activate and configure the Repeat and Reject Analysis feature.
Instructions are given in the Senographe DS Operator Manual.
Refer to the Browser chapter and locate the section Tools menu utilities - Medical Application prefer-
ences - Quality Check.
JC-ELE-A-067.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A067 - Repeat and Reject Analysis Configuration
Page no. 791
Senographe DS
Chapter 7