7. At the back of the Power Line Filter Box, connect the Earth cable (6) to the earth connector (5) (8 mm
open-ended wrench). The Earth cable originates from the white Gantry AC Supply cable (
2381711).
5
8. Ensure that the white Gantry AC Supply cable (
cable ties to secure it to the HV Unit (8) and the cables near the Power Line Filter Box (9).
9
JC-PHY-A-030.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A030 - Generator and Gantry Disconnection Reconnection
6
labeled 2381711)
Page no. 495
(7) is routed as shown below. Use
8
7