GE Senographe DS Manual page 835

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Code
Message text
006/029
Wrong value of sig-
nal
ADS_CART_ON_S
TATUS
007/xxx
E01 Generator fail-
ure
011/001
011/002
012/001
E02 Focus bias fail-
ure
012/002
E02 Focus bias fail-
ure
012/003
E02 Focus bias fail-
ure
012/004
E02 Focus bias fail-
ure
016/001
016/002
017/001
E03 Rotor failure
017/002
E03 Rotor failure
017/003
E03 Rotor failure
ERR-SUB-A-001.fm
Service Information and Procedures Class A 2385072-16-8EN
Description
At system power up, the Generator
CPU reports the Main Distribution
Board has detected the power at the
Control Station is not as expected (sig-
nal ADS_CART_ON_STATUS is not at
yes).
Possible causes are:
• the Control Station is not powered ON
• bad connection between J9 connector
on the Main Distribution Board and the
XJ11 connector on the Generator
Interface Board 400PL2.
• defective Main Distribution Board
Gene init tests: one fatal security input
<> 0; xxx identifies input. Access gener-
ator error log for full error decoding.
Focus bias software error: unknown
command
Focus bias software error: unknown
task status
Focus bias: signal Pres_Polar
(Bias_Presence) 'high' in standby sta-
tus.
Focus bias: signal Pres_Polar
(Bias_Presence) stays 'low' when bias
generation starts.
Focus bias: signal Pres_Polar
(Bias_Presence) goes 'low' after stabili-
zation.
Focus bias: signal Pres_Polar
(Bias_Presence) stays 'high' after bias
generation stopped.
Anode starter software error: unknown
command
Anode starter software error: unknown
task status
Anode starter: signal Pres_Rotation
(Rotation_Presence) 'high' in standby
status
Anode starter: signal Pres_Rotation
(Rotation_Presence) not rising during
acceleration to 9000 rpm
Anode starter: signal Pres_Rotation
(Rotation_Presence) not rising during
acceleration to 6000 rpm
Page no. 835
ERR SUB A001 - Generator Error Messages
Recommended actions
Proceed as follows:
• check that the Control Station is
powered ON
• if the problem persists, check the
cable connections between the J9
connector on the Main Distribution
Board and the XJ11 connector on
the Generator Interface Board
400PL2.
• if the problem persists, change the
Main Distribution Board
Senographe DS
Chapter 9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents

Save PDF