GE Senographe DS Manual page 312

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Cable Lay-out and Pin-out
PL301/J3-7
PL301/J3-8
PL301/J3-9
PL301/J3-10
PL301/J3-11
PL301/J3-12
1-59. W312 - Stepper Motor to Bucky Board
Stepper Motor M302 to Bucky Board PL302 J3:
OUT_1A
OUT_1B
OUT_2A
OUT_2B
1-60. W313 - Compression Board to Bucky Board
Compression Board PL303 J9 to connector J4a, then from J2a to Bucky Board PL302 J2:
PIN (PL303)
PL301/J9-1
PL301/J9-2
PL301/J9-3
PL301/J9-4
Chapter 3
Service Information and Procedures Class A 2385072-16-8EN
H-RT line 2
L-RT line 2
Not used
GND
Not used
Not used
Designation
Type
Hbridge
Hbridge
Hbridge
Hbridge
Designation
GND
Rec 0
Signal 1
Signal 2
RT-CAN
Diff 5 / 2.5 V
RT-CAN
Diff 2.5 / 0V
Ground
Level
0 / 48 Volts
0 / 48 Volts
0 / 48 Volts
0 / 48 Volts
Colour
Red
Black
Yellow
Green
Page no. 312
Senographe DS
PL302/J1-7
PL302/J1-8
0 V
PL302/J1-10
PIN
J3-1
J3-2
J3-3
J3-4
PIN (connector J4a)
PL302/J4a-6
PL302/J4a-3
PL302/J4a-4
PL302/J4a-5
Cables.FM

Hide quick links:

Advertisement

Table of Contents
loading