ERR SUB A001 - Generator Error Messages
Code
Message text
004/013
004/014
004/015
004/016
004/017
004/018
006/000
006/001
006/002
006/003
006/004
006/005
006/011
006/028
E20 Check Main
Distribution Rack
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Description
Gene interface: interrupt error on PTM1
timer 1
Gene interface: interrupt error on PTM1
timer 2
Gene interface: interrupt error on PTM1
timer 3
Gene interface: interrupt error on PTM2
timer 1
Gene interface: interrupt error on PTM2
timer 2
Gene interface: interrupt error on PTM2
timer 3 D208
Generator: fault at exposure command
Generator: fault at exposure command
Generator: signal kV_Inf_5
(Less_than_5kV) not rising
Generator: Signal En_Pose
(Exposure_On) not rising
Generator: fatal security triggered dur-
ing exposure
Generator: fatal security triggered at
end of exposure or signal En_Pose
(Exposure_On) stays 'high'
Generator: signal Secu_Rattrap
(No_Fatal_Security) stays 'high' after
'arcing recovery'
The Mains Distribution Rack is not cor-
rectly powered ON. The Low Voltage
Power Supply (LVPS) output at the J6
connector is too low to supply the Main
Distribution Assembly (expected volt-
age is 12 Vdc).
Page no. 834
Senographe DS
Recommended actions
Proceed as follows:
• Reboot the entire system.
• If the problem persists, check volt-
age at LVPS output J6 connector,
and adjust to 12 Vdc if required or
replace the LVPS.
• Reboot the Gantry and check no
error code displays on the X-ray
Console.
ERR-SUB-A-001.fm

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Subscribe to Our Youtube Channel

Table of Contents

Save PDF