GE Healthcare
Revision 1
2.
IEC60601-1-2 ELECTROMAGNETIC STANDARDS COMPLIANCE
2-1. General
This equipment complies with the IEC60601-1-2 Edition 2 EMC standard for medical devices.
The Senographe Equipment or System is suitable for use in electromagnetic environments as defined in
the limits and recommendations given in the following tables:
•
Emission Compliance level and limits (Table 1).
•
Immunity Compliance levels and recommendations for ensuring that the equipment retains its clinical
utility (Tables 2, 3, and 4).
Note:
This equipment complies with the above EMC standard when used with cables supplied by the
manufacturer up to the maximum lengths permitted by the system design specifications.
2-2.
Electromagnetic Emission
The Senographe system is suitable for use in the specified electromagnetic environment. The purchaser or
Operator of a Senographe system must ensure that it is used in an electromagnetic environment as
TABLE 1 - ELECTROMAGNETIC EMISSION
Emissions Test
Radio-Frequency
Emissions
CISPR11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Chap-Pre-Requirements.fm
Service Information and Procedures Class A 2385072-16-8EN
described below:
Compliance
The Senographe is primarily intended for use in non-domes-
Group1 Class A
tic environments, and not connected directly to the public
limits
mains supply network.
It is primarily intended for use in environments (such as hos-
pitals) with a dedicated supply system, and in an X-ray
shielded room.
The Senographe uses RF energy only for its internal func-
Group1 Class A
tion. The RF emission is therefore very low, and not likely to
limits
cause any interference in nearby electronic equipment.
Not applicable
The Senographe is primarily intended for use in non-domes-
tic environments, and not connected directly to the public
mains supply network.
Not applicable
The Senographe is primarily intended for use in non-domes-
tic environments, and not connected directly to the public
mains supply network.
Page no. 327
Pre-Installation System Requirements
Electromagnetic Environment
Senographe DS
Chapter 4
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated