GE Senographe DS Manual page 555

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
7-2-3
Enable the DICOM-1 Mode and Disable All Other Modes
So the Operator cannot select other display modes, only the 1-DICOM mode must be enabled.
1. Press the Enter button to display the Screen Manager window.
2. On the Screen Manager window that appears, leave the <Setup> item selected and press the Enter
button.
3. On the Setup window that appears, leave the <Mode Preset> item selected and press the Enter but-
ton.
4. On the Setup - Mode Preset window that appears, use the Up/Down navigation menu buttons to
select each mode, and the Left/Right navigation buttons to change the state to On/Off, as follows:
-
1 - DICOM
-
2 - Custom
-
3 - CAL
-
4 - Text
5. When the settings for each mode are as described and shown above, double tap the Enter button to
store the settings and close the Setup - Mode Preset window.
JC-ELE-A-008.fm
Service Information and Procedures Class A 2385072-16-8EN
On
Off
Off
Off
Page no. 555
Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
Senographe DS