GE Healthcare
Revision 1
ILLUSTRATION 7 - TROUBLESHOOTING FLOWCHART FOR LIFT MOVEMENT PROBLEMS RELATED TO
ERRORS 102/013 AND 102/014 - TYPE 3 LIFT SCREW ASSEMBLY
Perform Lift Functional Checks.
See
section 3-1
,
Lift Functional
Checks on page
1048.
NOK
Grease Checks
See
section
3-2
,
Grease
Checks on page
1048.
OK
Lift Gas Spring Checks
See
section 3-4
,
Lift Gas Spring
Checks on page
1052.
OK
Nut, Bearing, and Lift Checks
See
section 3-5
,
Bearing and
Lift Checks on page
1057.
OK
Check Lift Motor
See
section 3-6
,
Lift Motor
Checks on page
1057.
Note: If you have just changed
the Lift motor as instructed
above in illustration 6, the Lift
Motor can be assumed to be
OK, so proceed directly to
changing the Lift Board.
JC-TSG-A-015.fm
Service Information and Procedures Class A 2385072-16-8EN
Corrective actions
Grease Lift Screw
See
section 3-3
NOK
Screw on page
Change Lift Gas Spring
See
Job Card D/R A206A - Lift
NOK
Gas Spring on page
If Nut, Bearing or Lift NOK,
change Type 3 Lift without Mo-
tor (5373048), keeping the mo-
tor from the old lift.
NOK
See
Job Card D/R A232 - Lift
Screw Assembly on page
Change Type 3 Lift Motor
(5413527)
See
Job Card D/R A205 - Lift
Motor on page 1179
NOK
Change Lift Board
See
Job Card D/R A201 - Lift
OK
Board on page
Contact OLC.
Page no. 1047
Job Card TSG A015 - Lift Movement Troubleshooting
Perform Lift Functional Checks.
,
Grease Lift
See
1049.
Checks on page
Perform Lift Functional Checks.
See
1183.
Checks on page
Perform Lift Functional Checks.
See
Checks on page
1255.
Perform Lift Functional Checks.
See
.
Checks on page
Perform Lift Functional Checks.
1161.
See
Checks on page
Senographe DS
section 3-1
,
Lift Functional
1048.
NOK
section 3-1
,
Lift Functional
1048.
NOK
section 3-1
,
Lift Functional
1048.
NOK
section 3-1
,
Lift Functional
1048.
NOK
section 3-1
,
Lift Functional
1048.
NOK
OK
OK
OK
OK
OK
Chapter 9
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?