Job Card D/R A206A - Lift Gas Spring

1
SUPPLIES
Lift Gas Spring
2
TOOLS
Standard Tool Box
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 60 minutes
4
SAFETY PRECAUTIONS
WARNING
The lift gas spring has a larger mechanical energy when the arm is at heights of less than
the maximum height. To minimize the risk of rapid gas spring movement due to stored
mechanical energy, you must ensure that the Arm is raised to its maximum height before
you proceed.
WARNING
The Gantry must be powered off.
5
PREREQUISITES
The arm must be in the vertical position.
Raise the arm to its maximum height.
Switch off the Gantry power.
Remove the Front and Column covers for access (refer to
Gantry Covers on page