GE Senographe DS Manual page 1165

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
7
COMPLETION
7-1
Firmware Check
Check whether the new Lift Board firmware needs to be updated, as follows:
1. Reboot the Senographe system.
2. Go to the Service Desktop home page (see
3. Consult Table 1 on page 63 in the section
on page
63. Determine the system level from the ADS/IDC/Generator CPU versions displayed on the
Service Desktop home page, and by consulting Table 1 on page 63. The system level will be defined
by one of the System Level (Model Number) columns in Table 1 on page 63.
4. Check whether the current firmware version(s) for the new Lift Board corresponds to the required
firmware version for the that system level.
If the current firmware version(s) match the required version(s) quoted in Table 1 on page 63, no
additional steps are required.
If the current firmware version(s) does not match the required version(s), you must perform a firm-
ware update of the mismatched firmware version(s) so that the new Lift Board firmware version(s)
match those quoted in Table 1 on page 63. To update the new Lift Board firmware follow the
instructions supplied in the Gantry Software Update CD-ROM that is kept with the system.
JC-DR-A-201.fm
Service Information and Procedures Class A 2385072-16-8EN
Launching the Service Desktop on page
Component FW/SW to ADS Version compatibility Matrix
Page no. 1165
Senographe DS
Job Card D/R A201 - Lift Board
120).
Chapter 9

Hide quick links: