GE Healthcare
Revision 1
Job Card D/R A215 - Arm Control Keypad Cables
6-2
Removal and Reinstallation of Cables from the Arm Keypads
1. Switch off all Senographe electrical power.
2. Remove the Arm covers (see
3. Disconnect the keypad cable from the keypad and from connector J7 or J8 on the Compression
Board PL303.
4. Remove the cable.
5. Install the replacement cable.
6. Ensure that the keypad cables are positioned as shown:
Cable from right keypad is secured to the
Arm Distribution Board spacer as shown.
View from right, arm at 0°.
7
COMPLETION
Check for correct operation of all keypad controls.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Arm Covers and Handle on page
Keypad cables are connected
to the Compression Board.
Cable from left keypad
Cable from right keypad
Page no. 1218
526).
View from left, arm at 0°.
Senographe DS
JC-DR-A-215.fm