Job Card PHY A011 - Trackball Option Installation
cone mask in a clean place.
5. Carefully pull on the lower silicone mask (2) to remove it from the rotative arm support. Store the
lower silicone mask in a clean place.
6. Disconnect the mouse cable from the ADS computer.
7. Cut the cable ties that secure the monitor cables and mouse cable to the Control Station framework
Chapter 6
Service Information and Procedures Class A 2385072-16-8EN
Page no. 444
Senographe DS
1
2
JC-PHY-A-011.fm