GE Senographe DS Manual page 910

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
ERR SUB A003 - Gantry Error Messages
3-7
Error Codes from 160/yyy: CPU
Code
Category
160/000
HALT
160/001
HALT
160/002
HALT
160/003
HALT
160/004
HALT
160/005
HALT
160/006
HALT
160/007
HALT
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Text in Log File or
on LCD
INTERNAL SOFT-
One of the following out of memory errors occurred:
WARE ERROR
- functional device creation fails
- configuration interface creation fails
- HAL device creation fails
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
INTERNAL SOFT-
A Common Object Request Broker Architecture
WARE ERROR
(CORBA) extraction failed.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
INTERNAL SOFT-
An unknown attribute change notification, HAL sig-
WARE ERROR
nal, or Atlas event was received.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
INTERNAL SOFT-
The Gantry CPU failed to read the configuration file.
WARE ERROR
1. Reboot the system from the X-ray Console.
2. If the problem persists change the CPU Board
3. If the problem persists, contact your OLC.
INTERNAL SOFT-
The Gantry CPU failed to get the configuration data.
WARE ERROR
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
INTERNAL SOFT-
An Atlas event, attribute change, or HAL signal sub-
WARE ERROR
scription has failed.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
INTERNAL SOFT-
The Gantry CPU was unable to access the state
WARE ERROR
machine file.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
INTERNAL SOFT-
The Gantry CPU failed to get an attribute.
WARE ERROR
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Possible Cause/Recommended Action
and re-calibrate the Collimator afterwards.
Page no. 910
Senographe DS
Job Card
D/R A190
page
1131
CAL A029
page
1683
ELE A028
page 679
ERR-SUB-A-003.fm

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents