GE Healthcare
Revision 1
ERR SUB A003 - Gantry Error Messages
3-6
Error Codes from 150/yyy: PDU
Code
Category
152/000
DEBUG
ONLY
152/001
WARNING
152/002
DEBUG
ONLY
152/003
WARNING
152/004
DEBUG
ONLY
152/005
WARNING
152/006
DEBUG
ONLY
152/007
DEBUG
ONLY
152/008
DEBUG
ONLY
Chapter 9
Text in Log File or
on LCD
PDU 230V AC mains
is out of the spec lim-
its
12VDC Generator
voltage off
48V DC channel
error
Power Fail
Fan Fail
Over Temperature
Internal 12VDC volt-
age is out of spec
limits
Total current con-
sumption is too high
GND offset. Level of
PDU ground is too
high
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
The 230 V AC mains is out of the specified limits.
See
JC DIAG A1500 AC MAINS FAILURE on page
971
to fix the problem.
The 12 V DC signal from Generator Supply Com-
mand board to PDU board is not present.
See
JC DIAG A1501 12V DC Generator Voltage Off
on page 971
to fix the problem.
The AC/DC module is turned on but no 48 V DC is
detected.
See
JC DIAG A1502 ACDC MODULE POWER
FAILURE on page 973
The power on the AC/DC module has failed.
See
JC DIAG A1508 AC/DC Module Reports Power
Failure on page 976
to fix the problem.
The fan on the AC/DC module has failed.
See
JC DIAG A1509 AC/DC Module Reports Fan
Failure on page 976
to fix the problem.
The temperature of the AC/DC module has
exceeded the specified limits.
See
JC DIAG A1510 AC/DC Module Reports Over
Temperature on page 976
The internal (regulated) 12 V DC voltage is out of the
specified limits.
Ignore this error. If the error persists, contact your
OLC.
The total current consumption of PDU is too high (i.e.
greater than 40 A).
See
JC DIAG A1503 OVERCURRENT FAILURE on
page 974
to fix the problem.
The level of PDU ground is too high (possible noise
is present).
Check the Shunt R101 and cables W109, W111, and
W113.
If the problem persists, change PDU board (see
Card D/R A191 - PDU Board on page
Page no. 906
to fix the problem.
to fix the problem.
1137).
Senographe DS
Job Card
A1500
page 971
A1501
page 971
A1502
page 973
A1508
page 976
A1509
page 976
A1510
page 976
A1503
page 974
Job
ERR-SUB-A-003.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated