1-4. Operating Requirements - temperature and humidity
1-4-1.
Air conditioning
-
Air conditioning must be provided where necessary to ensure that no part of the equipment
(including the generator cabinet) operates in an ambient temperature exceeding 35°C (95ºF).
For patient comfort, ambient temperatures of 23°C ± 3°C (73ºF ± 5ºF) are recommended.
-
1-5. Short Term Shutdown Requirements - temperature and humidity
1-5-1.
Shutdown with Conditioner on
During evenings and weekends some Operators shutdown the Senographe system via the Control Sta-
tion; in that case the Conditioner remains switched on, and the environment must be within the following
limits.
Relative humidity (non-condensing)
1-5-2.
Shutdown with Conditioner off
During evenings and weekends some Operators shutdown the Senographe system via the circuit
breaker on the wall; in that case the Conditioner is switched off, and the environment must be within the
following limits.
Relative humidity (non-condensing)
1-5-3.
Air conditioning
Air conditioning must be provided where necessary to ensure that no part of the equipment (including the
generator cabinet) exists in an ambient temperature exceeding 35°C (95ºF) or below 15°C (59ºF).
Chap-Pre-Requirements.fm
Relative humidity (non-condensing)
Min.
10%
Min.
10%
Min.
5%
Service Information and Procedures Class A 2385072-16-8EN
Max.
80%
Max.
80%
15°C
Max.
Min.
95%
10°C
50ºF
Page no. 325
Pre-Installation System Requirements
Temperature
Min.
Max.
15°C
35°C
59ºF
95ºF
Temperature
Min.
Max.
35°C
59ºF
95ºF
Temperature
Max.
50°C
122ºF
Senographe DS