GE Healthcare
Revision 1
Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
7-1-3
Expanding the ADS GUI to Fill the Monitor (V3 and V4 Control Stations)
1. At the Login: prompt, login as sdc user to launch the Browser.
2. At the Browser login prompt, login as clinical user (password usually clinical).
Note:
At this stage the Browser will appear in the middle of the screen at a lower resolution than the
21-inch 3MP Monitor, with a black border around it. This is normal, and the following steps will con-
figure the ADS computer to use the full resolution of the 21-inch 3MP Monitor.
3. From behind the Browser, launch a command window (see
page
120).
A command window appears with an ADS prompt.
4. Change to the ~sdc/senovision/scripts directory, by entering:
cd ~sdc/senovision/scripts
5. Set the 3 MP Monitor option, by entering:
SetMonitorConfig.sh 3MP
The SetMonitorConfig.sh script runs, as follows:
## Running as sdc...
Run SetMonitorConfig.sh script with input SET 3MP
Enter root password:
Password:
Sun Microsystems Inc.
Run SetScreenConfig script with input 3MP
Remove existing setscreenenv file.
Creating symbolic link /export/home/sdc/senovision/scripts/setscreenenv -> /export/home/
sdc/senovision/scripts/setscreen
env3MP
Enter root password :
Password:
6. At the Password: prompt during the execution of the SetMonitorConfig.sh script, enter the root pass-
word (operator).
The SetMonitorConfig.sh script continues to run, and when complete displays the following mes-
sages:
Sun Microsystems Inc.
Current lang ext set
Current lang ext set
Current lang ext set
Sun Microsystems Inc.
END of Script SetScreenConf - PLEASE REBOOT
End of Script SetMonitorConfig - PLEASE REBOOT
7. Close the command line window, by entering:
exit
8. Shutdown the system from the Browser shutdown button.
9. Wait two minutes for the UPS to be ready, then power up the system from the X-ray Console.
Chapter 7
Service Information and Procedures Class A 2385072-16-8EN
SunOS 5.10
Generic January 2005
SunOS 5.10
Generic January 2005
SunOS 5.10
Generic January 2005
Page no. 552
Launching a Command Line Window on
Senographe DS
JC-ELE-A-008.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated