GE Senographe DS Manual page 314

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Cable Lay-out and Pin-out
1-62. W315 - 48v dc-LP2
PDU Board PL101 J6 to Arm Distribution Board PL301 J2:
1-63. W316 - Image Receptor Earth Connect
Ground bar G301 to Image receptor.
1-64. W317 - Compression Trolley Earth Connect
Ground bar G301 to Compression Trolley.
1-65. W318 - Compression Block Connect
Manual Brake B301 to Paddle Detect. Board PL304:
Green/Blue (-) Green/Red (+)
1-66. W319 - Force Sensor
Force Sensor B302 to Paddle Detect. Board PL304.
1-67. W320 - Earth Cable for Paddle Holder
Paddle Holder PL304 to Ground.
1-68. W401 - 48v dc-LP2
Arm Distribution Board PL301 J8 to Collimator DC/DC PL402 J5:
1-69. W402 - Thermoswitch Connect
Chapter 3
Service Information and Procedures Class A 2385072-16-8EN
N° Pin
J6-1
J6-2
PIN
J1-1
J1-2
Designation
Power supply
secu_sheath
Functions
+ 48 Volts
GND
Function
0 Volts
+ 48 Volts
Level
15 V
15 V or 0 V(failure)
Page no. 314
N° Pin
J2-1
J2-2
PIN
J8-2
J8-1
Pin N°
J4-2
J4-1
Senographe DS
Cables.FM

Hide quick links:

Advertisement

Table of Contents
loading