Job Card D/R A227 - Rotation Potentiometer
6-4
Troubleshooting
If the error message 112/001 appears after potentiometer replacement, check the voltage between
the center pin of the rotation potentiometer connector and ground.
The following values must be obtained, according to the arm position:
-
Arm at +180°;
-
Arm at 0°;
-
Arm at -165°;
7
COMPLETION
Carry out the following calibration:
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
V = 2.825 V ±0.05 V dc.
V = 1.65 V ±0.05 V dc.
V = 0.606 V ±0.05 V dc.
Job Card CAL A021 - Arm Rotation Calibration on page
Page no. 1242
Senographe DS
1679.
JC-DR-A-227.fm