Job Card ELE A028 - Collimation Checks

1
OBJECTIVE
To ensure that:
there is no excessive extension of the X-ray field beyond the edges of the image receptor
the X-ray field aligns with the light field,
the chest wall edge of the compression paddle aligns with the chest wall edge of the image receptor
2
SUPPLIES
None
3
TOOLS
One coin (e.g., a nickel, or a 5-Eurocent coin).
4 strips of
GAFCHROMIC® XR-M
Aluminium attenuator plate
Radiation Protection Plate
4
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 20 minutes
5
SAFETY PRECAUTIONS
The procedures in this Job Card produce X-rays. Ensure that you take appropriate precautions when
operating with X-rays (for example, stay behind the radiation screen during an X-ray exposure).
6
PREREQUISITES
None