2. The X-ray field must extend beyond all edges of the active image receptor. The X-ray field must not
extend beyond any edge of the image receptor by more than that quoted in
For each of the exposure configurations below, enter the determined deviations between the edge of
the X-ray field and the edge of the image receptor in the plane of the image receptor for each of the
four XR-M film strips. That is, the value Z
2-2.
For each of the exposure configurations below, determine whether the action limit for each edge has
passed or failed according to the table below (see illustration 1 on
TABLE 1 - X-RAY FIELD - IMAGE RECEPTOR ACTION LIMITS
Film
A
D
B
Posterior (chest wall)
C
Mo/Mo
Film
A
D
B
Posterior (chest wall)
C
Rh/Rh
Film
A
D
B
Posterior (chest wall)
C
JC-ELE-A-028.fm
Edge
Left
• If Z
required.
Right
• If Z
Anterior
Consult the physicist regarding the local requirements, and if the
test has failed re-calibrate the Collimator according to
CAL A029 - Collimator Format Calibration on page
perform the Collimator Checks.
• If Z
mator according to
bration on page
• If Z
ther action required.
• If Z
test may have failed. Consult the physicist regarding the local
requirements, and if the test has failed re-calibrate the Collimator
according to
on page
• If Z
tor according to
tion on page
Edge
Left
Right
Anterior
Edge
Left
Right
Anterior
Service Information and Procedures Class A 2385072-16-8EN
for each XR-M film strip determined in step
i
is below 10 mm, the test has passed, and no further action
i
is between 10 mm and 13.2 mm, the test may have failed.
i
is above 13.2 mm, the test has failed. Re-calibrate the Colli-
i
Job Card CAL A029 - Collimator Format Cali-
1683, then re-perform the Collimator Checks.
is between 4 mm and 7 mm, the test has passed, and no fur-
i
is between 0 mm and 4 mm or between 7 mm and 10 mm, the
i
Job Card CAL A029 - Collimator Format Calibration
1683, then re-perform the Collimator Checks.
is above 10 mm, the test has failed. Re-calibrate the Collima-
i
Job Card CAL A029 - Collimator Format Calibra-
1683, then re-perform the Collimator Checks.
Pass / Fail / Comment (see
Z
i
Z
Pass / Fail / Comment (see
i
Page no. 691
Job Card ELE A028 - Collimation Checks
Table
page 681
for film positions):
Action Limit
Senographe DS
1.
4
of
section 7-
Job Card
1683, then re-
Table
1)
Table
1)
Chapter 7