Revision 1
Job Card ELE A021 - Networking Connection and Configuration
18. At the Confirm Information screen that appears displaying the date and time settings, press the <F2>
key to continue.
Confirm Information
> Confirm the following information. If it is correct ...
.........................
Time zone:
Date and time: 2007-06-09 11:33
19. The Root Password screen appears prompting you to specify a root password for the Acquisition
Workstation.
Root Password
Please enter the root password for this system.
The root password may contain alphanumeric and special characters. For
security, the password will not be displayed on the screen as you type it.
> If you do not want a root password leave both entries blank.
Root password:
Root password:
Type the required root password (usually operator) in both the Root password fields and press the
<Return> key when done. The password you type is not displayed. The default password for root is
operator.
Note:
Ensure that the caps-lock key is not engaged before entering the password.
If there is a difference between the two entries, the system will request it again.
Press the <F2> key to continue.
20. A System identification screen appears.
System identification is completed.
This system is configured with NFS version 4, which uses a domain
name that is automatically derived from the system's name services.
The derived domain name is sufficient for most configurations. In a
few cases, mounts that cross different domains might cause files to
be owned by 'nobody' due to the lack of a common domain name.
Do you need to override the system's default NFS version 4 domain
name (yes/no) ? [no] : _
At the prompt, press the <Return> key to accept the default answer (no).
The system reboots. Log in to the system according to
Chapter 7
Service Information and Procedures Class A 2385072-16-8EN
 
 
GMTn
Page no. 642
           
     
 
 
 
           
 
 
 
 
 
 
 
 
 
 
Login on page
Senographe DS
     
           
     
652.
JC-ELE-A-021.fm

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.